InTemp Bluetooth Low Energy Temperature (with Glycol) Data Logger
Description
InTemp Bluetooth Low Energy Temperature (with Glycol) Data Logger InTemp ® CX402-TXXX (with Glycol Bottle) data loggers monitor
temperatures in refrigerators, freezers and other temperature-controlled
environments to compile with global GDP requirements. The loggers
communicate wirelessly via Bluetooth Low Energy to mobile devices. Using
the InTemp app, you can easily view data, check logger status, set
alarms, and create and share secure PDF reports for streamlined
reporting and regulatory compliance. All loggers feature a built-in LCD
display to view current and daily min and max temperatures, advanced
audible and visual alarm capabilities for notification of temperature
excursions, an internal sensor for ambient temperature monitoring, and a
single point NIST Certification of Calibration for both the logger body
and probe.
Models are available with a 2- or 4-meter external temperature probe
attached to a Glycol bottle used to insulate the probe and delay thermal
response times.
Feature
- Easy-to-read display with current and daily min/max temperatures
- Audible and visual alarms
- Bluetooth transmission of data and alarm states
- Logger setup and download of data in seconds – without a PC or cable
- +/- 0.5C accuracy for reliable temperature readings
- Battery-powered to guard against local power loss
- Several Glycol bottle/probe-only options available
- Optional ISO 17025 compliant calibration service (NIST included)
Spesification
To see the full specifications for this product, please see our product manual found under the documentation tab if available.
Logger
|
|
Radio Power
|
1 mW (0 dBm)
|
Transmission Range
|
Approximately 30.5 m (100 ft) line-of-sight
|
Wireless Data Standard
|
Bluetooth Low Energy (Bluetooth Smart)
|
Logger Operating Range
|
-30° to 70°C (-22° to 158°F), 0 to 95% RH (non-condensing)
|
Logging Rate
|
1 second to 18 hours
|
Time Accuracy
|
±1 minute per month at 25°C (77°F)
|
Battery Type
|
Two AAA 1.5 V alkaline or lithium batteries, user replaceable
|
Battery Life
|
1 year, typical with logging interval of 1 minute. Faster
logging intervals, remaining connected with the InTemp app, excessive
report generation, numerous audible alarms, and paging all impact
battery life.
|
Memory
|
128 KB (84,650 measurements, maximum)
|
Full Memory Download Time
|
Approximately 60 seconds; may take longer the farther the device is from the logger
|
LCD
|
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range
|
Dimensions
|
9.4 x 4.5 x 2.59 cm (3.7 x 1.77 x 1.02 inches)
|
Weight
|
90.2 g (3.18 oz)
|
Environmental Rating
|
IP54
|
CE
|
The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).
|
External Temperature Probe
|
|
Range
|
-40° to 100°C (-40° to 212°F)
|
Accuracy
|
±0.5°C from -15° to 80°C (±0.9°F from 5° to 176°F)
±1.0°C from -40° to -15°C (±1.8°F from -40° to 5°F) ±0.7°C from 80° to 100°C (±1.3°F from 176° to 212°F) |
Resolution
|
0.024°C at 25°C (0.04°F at 77°F)
|
Drift
|
<0.1°C (0.18°F) per year
|
Response Time
|
With 5 mL bottle: 10 minutes typical to 90% in air moving 1 m/s
With 15 mL bottle: 20 minutes typical to 90% in air moving 1 m/s With 30 mL bottle: 25 minutes typical to 90% in air moving 1 m/s |
Cable Length
|
2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable
|
Ambient Temperature Sensor
|
|
Range
|
-30° to 70°C (-22° to 158°F)
|
Accuracy
|
±0.5°C from -15° to 70°C (±0.9°F from 5° to 158°F)
±1.0°C from -30° to -15°C (±1.8°F from -22° to 5°F) |
Resolution
|
0.024°C at 25°C (0.04°F at 77°F)
|
Drift
|
<0.1°C (0.18°F) per year
|
Response Time
|
15 minutes typical to 90% in air moving 1 m/s
|